The high performance Heat-flow Reaction Calorimeter
Suitable for non-isothermal Heat-flow Reaction Calorimetry
|
|
|
|
|
|
 |
 |
High performance calorimeter, non-isothermal advanced type
(n-iso HFC)
|
 |
Automated „Zero Watt“ baseline function, instrument supported evaluation |
 |
Measurement can be performed without simultaneous calibration |
 |
Automatic cp determination during calibration |
 |
On-line display of output, heat, A, U and cp |
|
 |
Optional heat-balance within reflux condenser |
|
| Measuring principles & Blue Window |

HF = (TR - TJout) * A * U
(W)......(°C)..(°C).....(m2)...(W/m2.K)
|

|
|
|
|
|
| Data Summary Calo 2310 base |
Evaluation Calo 2310 base |
| Reactor |
Glass, cylindrical, 1l capacity, double jacket, with bottom discharge valve, sealed, with safety valve |
| Operating Range |
-20°C to +180°C |
| Thermostat |
Unistat Tango from -40°C to +200°C |
| Stirrer Speed |
40 to 2000 rpm |
| Dosing Mode |
1 x gravimetrically, using membrane pump |
| Pressure/Vacuum |
0 to 1.6 barabs, including control |
| Temp.-Resolution |
0.007 K, HF signal 0.001 K |
| Accuracy |
3% (isothermal, 5° to 50°C) |
| OPTIONS |
Heat balance reflux (accuracy 7%) second dosing facility, pH, pH control, distillation with balance, pressure reactor version |
|
 |
| SysGraph evaluation software combined with Calo 2000 package for automated „Zero Watt“ base-line function, output and heat of HF and optional reflux as well as adiabatic increase TR. |
|
| Download more Information |
| Short Overview |
Detailed Leaflet |
The Standards with Principles and FAQ's |
Calo 2310 Flyer
File size 384 kB
4 pages A4, English
|
Calo 2310 eco-base-pro
File size 1'736 kB
28 pages A4, English
|
-->
The main paper about Reaction Calorimetry
Takes some time for download
|

White Paper
File size 3'697 kB
52 pages A4, English
|
|
|
|
|
|
|
|
|
|
|
|
|